Skip to content
New issue

Have a question about this project? Sign up for a free GitHub account to open an issue and contact its maintainers and the community.

By clicking “Sign up for GitHub”, you agree to our terms of service and privacy statement. We’ll occasionally send you account related emails.

Already on GitHub? Sign in to your account

R18 must be connected to RLD #19

Open
vredzhepov opened this issue Jul 16, 2019 · 2 comments
Open

R18 must be connected to RLD #19

vredzhepov opened this issue Jul 16, 2019 · 2 comments

Comments

@vredzhepov
Copy link

R18 must be connected to RLD instead of SW

@selman-nus
Copy link

@vredzhepov i am very late but i agree. considering CARDIAC MONITOR CONFIGURATION on pg 25 of the AD8232 datasheet this design is correct. However R18 seems to be designed as a optional jumper and when it is selected design should serve as HEART RATE MEASUREMENT NEXT TO THE HEART configuration as on pg 24. So in summary it should be connected to RLD.

If you are not going to change the configuration, you are safe to use this design without any problems tho..

@vredzhepov
Copy link
Author

@selman-nus I think that this bug must be denoted in the official documentation for the board.

Sign up for free to join this conversation on GitHub. Already have an account? Sign in to comment
Labels
None yet
Projects
None yet
Development

No branches or pull requests

2 participants